Product Description
TECHNICAL SPECIFICATIONS
FTBx-740C-CWDM FTBx-740C-DWC
Laser nominal wavelength (nm)
1270, 1290, 1310, 1330, 1350, 1370, 1390, 1410,
1430, 1450, 1470, 1490, 1510, 1530, 1550, 1570,
1590, 1610
C-Band tunable 1527.99-1567.95 nm
ITU-T G694.1 Channels 12-62
(191.2 THz - 196.2 THz)
Central wavelength uncertainty (nm) a ±3 DWDM 50Ghz channel wavelength control
Channel spacing tuning N/A 50 GHz and 100 GHz increments on
ITU-T G694.1 grid
Dynamic range at 20 µs (dB) b >37 40
Event dead zone (m) c 1.1 0.7
Attenuation dead zone (m) c 5 3.5
Distance range (km) 0.1 to 400 0.1 to 400
Pulse widths (ns) 5 to 20 000 5 to 20 000
Sampling points Up to 256 000 Up to 256 000
Sampling resolution (m) 0.04 to 10 0.04 to 10
Distance accuracy (m) d ±(0.75 + 0.0025 % x distance + resolution) ±(0.75 + 0.0025 % x distance + resolution)
LASER SAFETY
GENERAL SPECIFICATIONS
Size (H x W x D) 158 mm x 24 mm x 174 mm (6 ¼ in x 15/16 in x 6 7/8 in)
Weight 0.4 kg (0.9 lb)
Temperature Operating
Storage
Refer to platform’s specification sheet
–40 °C to 70 °C (–40 °F to 158 °F)
Relative humidity 0% to 95% non-condensing
FTBx-740C-CWDM FTBx-740C-DWC
Laser nominal wavelength (nm)
1270, 1290, 1310, 1330, 1350, 1370, 1390, 1410,
1430, 1450, 1470, 1490, 1510, 1530, 1550, 1570,
1590, 1610
C-Band tunable 1527.99-1567.95 nm
ITU-T G694.1 Channels 12-62
(191.2 THz - 196.2 THz)
Central wavelength uncertainty (nm) a ±3 DWDM 50Ghz channel wavelength control
Channel spacing tuning N/A 50 GHz and 100 GHz increments on
ITU-T G694.1 grid
Dynamic range at 20 µs (dB) b >37 40
Event dead zone (m) c 1.1 0.7
Attenuation dead zone (m) c 5 3.5
Distance range (km) 0.1 to 400 0.1 to 400
Pulse widths (ns) 5 to 20 000 5 to 20 000
Sampling points Up to 256 000 Up to 256 000
Sampling resolution (m) 0.04 to 10 0.04 to 10
Distance accuracy (m) d ±(0.75 + 0.0025 % x distance + resolution) ±(0.75 + 0.0025 % x distance + resolution)
LASER SAFETY
GENERAL SPECIFICATIONS
Size (H x W x D) 158 mm x 24 mm x 174 mm (6 ¼ in x 15/16 in x 6 7/8 in)
Weight 0.4 kg (0.9 lb)
Temperature Operating
Storage
Refer to platform’s specification sheet
–40 °C to 70 °C (–40 °F to 158 °F)
Relative humidity 0% to 95% non-condensing